Series 46 Cryogenic Temperature Probes

Discontinued, Please See Our New Series 56 Probes

Scientific Instruments offers a variety of temperature probes that allow mounting in hazardous environments. These units are constructed of stainless steel, hermetically sealed, and the sensing element is mounted in a thermal epoxy at the tip. The probes are designed for direct immersion or thermowell applications.

The basic sensing elements used in the Series 46 temperature probe assemblies, designed and manufactured by Scientific Instruments, is the Si410 silicon diodes. These diodes are rugged, wide-range (1.5K to 450K) sensing elements which offer excellent interchangeability to standard temperature-versus-voltage curves. This interchangeability allows a user to employ any given temperature probe assembly in his facility while using a single temperature curve for signal conditioning requirements. This model has been used for ground support and flight hardware. See below for detailed specifications.

SKU: SI-Series46 Category:

Additional information

Enclosure

All wetted areas are 316 SS.

Insertion Length

Customer specified.

Sensing Element

The sensing element is a silicon diode supported and contained in a closed housing.

Media Connection

Boss seal per MS33656E4

Temperature Range

Each unit is capable of operating within an environmental temperature range of 1.5 K (-271.66°C) to 400 K (126.85°C)

Operating Current

The operating current of the unit should be 10 microamperes (µA) ± 0.1µA for the 46D units and 100µA ± 1µA for 46DX and 46DX1 units (the sensor excitation is supplied by an external signal conditioner).

Service Media

The unit will operate in the following gaseous or liquid media where compatible with the sheath-housing material and within the temperature range of the unit: Alcohol Neon (Ne) Ammonia (NH₃) Nitrogen (N₂) Helium (He) Nitrogen teroxide (N₂O₄) Hydrogen (H₂) Oxygen (O₂) Hydrazine (N₂H₄) Unsymmetrical dimethylhydrazine (UDMH) Monomenthylhygrazine (MMH) Water (H₂O)

Repeatability

The unit is capable of repeating test parameter results under identical test conditions, within ± 0.1K (± 5 mVDC) at the boiling point of liquid helium and within ± 0.5K (1.2 mVDC) from 25 K to 400 K.

Accuracy

The unit shall be accurate to within ± 0.25K (± 12.8 mVDC) from 1.5K to 25K and within ± 0.5K (1.2 mVDC) from 25K to 400K. (46DX and 46DX1, within ± 0.75K from 273K to 400K)

Response Time

The unit shall respond within 63.2 percent of total temperature change, for a step change in temperature from 293.15 K ± 3K to 77.35 K ± 3K, within 15 seconds.

Probe Type

Series 46D Aerospace Probe (OMEGA), 2.5 Inch Insertion Length, Series 46DX (100uA) Aerospace Probe, 1/8” OD, 12 Inch Insertion Length (+/- 1K from 20K to 77K), Series 46DX1 (100uA) Aerospace Probe, 2.2 Inch Insertion Length, Extended Pressure (5000 psi MEOP, Proof = 1.2X = 6000 psi, Burst = 2X = 10000 psi)